Sample Analysis Time:
2.5 minutes (Gases only)
4.0 minutes (Chemistry/Gas & Osmolality)
6.25 minutes (All 15 tests)

Operating Temperature Range:
15-30°C (59-86°F)

Operating Relative Humidity Range:
20-85%

Sample Size:
1 mL (for all modules)

Sampling Options:
Individual via syringe/cup
Automated batch using 20-position sampling tray
On-line, automated from up to 10 bioreactors

Host Computer Operating System:
Windows XP

Electrical Requirements:
90-264 VAC, 50 to 60 Hz (Universal Power Supply)

System Size:
Height: 20 in (50 cm)
Width: 26 in (66 cm)
Depth: 20 in (50 cm)

System Weight:
Analyzer: 94 lbs (42.6 kg) without reagents
Host Computer Monitor/CPU: 41.5 lbs (18.8 kg)

Safety Certifications:
TUV: IEC 61010-1:2001

Quality Systems Certifications:
TUV: ISO9001:2000 FDA Registered Quality Compliant, PAT Compatible, 21 CFR Part 11 Compliant

Chemistry/Gas Module
Assay Measurement Range Resolution Method
Glucose 0.2-30.0 g/L* 0.01 g/L Biosensor
Lactate 0.2-10.0 g/L* 0.01 g/L Biosensor
Glutamine 0.2-12.0 mmol/L* 0.01 mmol/L Biosensor
Glutamate 0.2-12.0 mmol/L* 0.01 mmol/L Biosensor
Ammonium 0.2-25.0 mmol/L 0.01 mmol/L Direct ISE
pH 5.000-8.000 0.001 Direct ISE
PCO2 3.0-300.0 mmHg 0.1 mmHg Direct ISE
PO2 3.0-800.0 mmHg 0.1 mmHg Clarke Electrode
Sodium 40-220 mmol/L 0.1 mmol/L Direct ISE
Potassium 1.0-25.0 mmol/L 0.01 mmol/L Direct ISE
Calcium 0.10-10.00 mmol/L 0.01 mmol/L Direct ISE

*Ranges reflect user selectable 1:2 dilution

Osmolality Module
Assay Measurement Range Resolution Method
Osmolality 0-1500 mOsm/kg 1 mOsm/kg Freezing Point

Cell Density/Viability Module
Assay Measurement Range Resolution Method
Cell Diameter 8-50 µm N/A Digital Imaging
Detectable Density 5x104-2x107 cells/mL** N/A Digital Imaging
Viability 0-100% N/A Digital Imaging

**Ranges reflect user selectable 1:2 dilution

Phosphate Module
Assay Measurement Range Resolution Method
Phosphate, under development 0.2-25.0 mmol/L 0.1 mmol/L Photometry

Calculated Tests: O2 Saturation; CO2 Saturation; Temp. Corrected pH, PCO2, PO2

Back to top