Sample Analysis Time: 2.5 minutes (Gases only) 4.0 minutes (Chemistry/Gas & Osmolality) 6.25 minutes (All 15 tests)
Operating Temperature Range: 15-30°C (59-86°F)
Operating Relative Humidity Range:
20-85%
Sample Size: 1 mL (for all modules)
Sampling Options: Individual via syringe/cup Automated batch using 20-position sampling tray
On-line, automated from up to 10 bioreactors
Host Computer Operating System: Windows XP
Electrical Requirements: 90-264 VAC, 50 to 60 Hz (Universal Power Supply)
System Size: Height: 20 in (50 cm) Width: 26 in (66 cm) Depth: 20 in (50 cm)
System Weight: Analyzer: 94 lbs (42.6 kg) without reagents Host Computer Monitor/CPU: 41.5 lbs (18.8 kg)
Safety Certifications: TUV: IEC 61010-1:2001
Quality Systems Certifications: TUV: ISO9001:2000 FDA Registered Quality Compliant, PAT Compatible, 21 CFR Part 11 Compliant
Chemistry/Gas Module
| Assay |
|
Measurement Range |
|
Resolution |
|
Method |
| Glucose |
|
0.2-30.0 g/L* |
|
0.01 g/L |
|
Biosensor |
| Lactate |
|
0.2-10.0 g/L* |
|
0.01 g/L |
|
Biosensor |
| Glutamine |
|
0.2-12.0 mmol/L* |
|
0.01 mmol/L |
|
Biosensor |
| Glutamate |
|
0.2-12.0 mmol/L* |
|
0.01 mmol/L |
|
Biosensor |
| Ammonium |
|
0.2-25.0 mmol/L |
|
0.01 mmol/L |
|
Direct ISE |
| pH |
|
5.000-8.000 |
|
0.001 |
|
Direct ISE |
| PCO2 |
|
3.0-300.0 mmHg |
|
0.1 mmHg |
|
Direct ISE |
| PO2 |
|
3.0-800.0 mmHg |
|
0.1 mmHg |
|
Clarke Electrode |
| Sodium |
|
40-220 mmol/L |
|
0.1 mmol/L |
|
Direct ISE |
| Potassium |
|
1.0-25.0 mmol/L |
|
0.01 mmol/L |
|
Direct ISE |
| Calcium |
|
0.10-10.00 mmol/L |
|
0.01 mmol/L |
|
Direct ISE |
Osmolality Module
| Assay |
|
Measurement Range |
|
Resolution |
|
Method |
| Osmolality |
|
0-1500 mOsm/kg |
|
1 mOsm/kg |
|
Freezing Point |
Cell Density/Viability Module
| Assay |
|
Measurement Range |
|
Resolution |
|
Method |
| Cell Diameter |
|
8-50 µm |
|
N/A |
|
Digital Imaging |
| Detectable Density |
|
5x104-2x107 cells/mL** |
|
N/A |
|
Digital Imaging |
| Viability |
|
0-100% |
|
N/A |
|
Digital Imaging |
Phosphate Module
| Assay |
|
Measurement Range |
|
Resolution |
|
Method |
| Phosphate, under development |
|
0.2-25.0 mmol/L |
|
0.1 mmol/L |
|
Photometry |
Calculated Tests: O2 Saturation; CO2 Saturation; Temp. Corrected pH, PCO2, PO2