Methodology:
Electrochemical, Lactate Oxidase Biosensor
Test Measured: Blood Lactate
Test Time: 13 seconds
Test Strip Volume: 0.7 µL
Battery Life (normal): 600 Tests
Data Output Port: Serial
Electrochemical Biosensor: Yes
Calculates Average Results: Yes
Data Storage: Users/Tests: 10/130
Data Transfer Port: Yes
Weight: 2.65 ounces (75g)
Size: 3.6" x 2.3" x 0.9" (91.4mm x 58.4mm x 22.9mm)
Measurement Range:
Lactate: 0.3 to 25.0 mmol/L
Operating Ranges:
Temperature: Lactate 41°F to 113°F (5°C to 45°C)
Altitude: 0 to 12,000 feet
Humidity: 10% to 90% relative humidity
On-board
Data Storage Files:
1 common file for 20 tests
9 additional files for 10 tests
each
2 Quality Control files for
20 tests
|
Typical
Analytical Specifications
for Imprecision: |
Within-Run
Imprecision n=20,
whichever is greater |
|
Day-to-Day
Imprecision n=20,
whichever is greater |
|
| Analyte |
|
CV%
|
|
SD |
|
CV% |
|
SD |
| Lactate |
|
5.0 |
|
0.2
(mmol/L) |
|
7.0 |
|
0.3
(mmol/L) |
|
Battery:
Coin Type No. 2450
Optional Accessories:
Meter to PC data cable, with USB Port Adapter and PC data charting program (Windows® 2000 and beyond)
Meter Warranty:
Warranty effective for one year from purchase date.
Certifications and Compliance:
EMC (EN 55022 class B, IEC 801-2, EN 61000-4-3, ENV 50204, EN 61000-4-2)
FOR PROFESSIONAL USE ONLY
|