 Critical Care Blood Gas Analyzer
Critical Care Blood Gas Analyzer Stat Profile Prime Plus is a comprehensive, whole blood critical care analyzer that combines blood gases, electrolytes, metabolites, CO-Oximetry, and 34 calculated results in a simple, compact analyzer. Prime Plus combines maintenance-free, component cartridge technology for sensors and reagents with patented, new, maintenance-free, and non-lysing whole blood CO-Oximetry technology.
Prime Plus results are produced rapidly, a complete test menu panel in about 90 seconds, and are combined with bidirectional connectivity, a robust data management system, and comprehensive cybersecurity protection.

		       Urea (BUN), Creatinine and eGFR
Urea (BUN), Creatinine and eGFR
				  Over 50% of patients admitted to the ICU will develop some stage of acute kidney injury (AKI).1  Stat Profile Prime Plus is the only blood gas analyzer  to provide  whole blood urea (BUN) and creatinine (plus eGFR) test options for rapid  assessment of kidney function.
              
                Ionized magnesium (iMg)
              Disruptions in the balance of iMg, Na, K, and iCa can cause cardiac  arrhythmias, reduced cardiac contraction, and cardiac arrest. Prime Plus  is the only blood gas analyzer  to provide a comprehensive profile of  electrolytes including iMg.
Estimated Plasma Volume (ePV) 
Prime Plus analyzers have the unique ability to report estimated plasma volume (ePV), a very important test that assesses the intravascular fluid content of blood, the patient’s hydration status. The plasma volume status of a patient is one of the top priorities in managing many different conditions including shock, sepsis, congestive heart failure, acute or chronic kidney disease, chronic pulmonary disease, as well as general postoperative care.
Prime Plus reports ePV and ΔPV using the Strauss equation which requires measured hemoglobin and measured hematocrit to determine ePV. ePV may be most beneficial when it is measured serially, and thus the change in plasma volume (ΔPV) can also be used to guide fluid therapy.
Prime Plus incorporates a new, patented multi-wavelength optical system that scans a continuous spectrum of optical wavelengths to enable a comprehensive CO-Oximetry panel result without lysing the sample. The optical components in contact with blood are contained in the disposable sensor card.
              CO-Oximetry test menu
O2Hb     COHb     MetHb     HHb      tHb   HbF*    tBil*
NovaNet bidirectional middleware for all Nova connected devices 
NovaNet ensures timely, accurate capture of Nova analyzer   test results  for clinicians and managers to retrieve wherever and whenever needed. Also included is comprehensive cybersecurity protection and  encryption that provides  protection  against  attempts to access a hospital's network. 

 Liquid QC provides the only reliable test of analyzer performance
Liquid QC provides the only reliable test of analyzer performance
        United States federal government regulations and many international government regulations have eliminated electronic equivalent QC and are requiring true
liquid QC.2
 Automated QC complies with U.S. CLIA, German RiLiBAK, and other international QC requirements
   QC cartridges contain a 30-day supply of liquid QC material. Controls  are run automatically at user-selected intervals. Prime Plus quality  controls:
 

Prime Plus  provides an automated electronic quality monitoring supplement to liquid QC. SQM continuously monitors the status and performance of all analytical components (including sensors, reagents, calibrators, sample integrity, software, and electronics), providing real-time, sample-to-sample assurance of correct performance.
   
 
    For more  product information, click here or contact Nova Biomedical at: 
				 Nova Biomedical / 200 Prospect Street / Waltham, MA 02454 / 781-894-0800 
				 For Nova Biomedical Sales inquiries, reach us at 1-800-458-5813.
For Technical Support, reach us at 1-800-545-NOVA (6682).
To place an order, call 1-800-822-0911 or email novaorders@novabio.com
1. Mandelbaum T et al. Outcome of critically ill patients with acute kidney injury using the Akin criteria. Crit Care Med 2011;39:2259-2264.
   2. Centers for Medicare and Medicard Services, Center for Clinical Standards and Quality/Survey and Certification Group. Policy clarification or acceptable control materials used when quality control (QC) is performed in laboratories. Baltimore, MD: CMS, April 8, 2016. 
        
| Critical Care Test Menu | Methodology | 
| pH | Direct ISE | 
| PCO2 | Severinghaus | 
| PO2 | Amperometric | 
| SO2% | Optical, reflectance | 
| Hematocrit | Conductivity/Na correction | 
| Na | Direct ISE | 
| K | Direct ISE | 
| Cl | Direct ISE | 
| iCa | Direct ISE | 
| iMg | Direct ISE | 
| Glucose | Enzyme/Amperometric | 
| Lactate | Enzyme/Amperometric | 
| Urea (BUN) | Enzyme/Amperometric | 
| Creat | Enzyme/Amperometric | 
Calculated Tests
| eGFR | A-aDO2 | ePV and delta PV | 
| HCO3- | a/A | Normalized iCa | 
| TCO2 | PO2/FIO2 | Normalized iMg | 
| BE-ecf | Anion Gap | Osmolality | 
| BE-b | SBC | Hemoglobin | 
| A | Base Excess | O2 Saturation | 
| ePV (Estimated Plasma Volume) | ||
| pH/PCO2/PO2 Corrected to Patient Temperature Respiratory Index (If % FIO2 value entered) Actual Bicarbonate Standard Bicarbonate Oxygen Index (OI) | ||
| CO-Oximetry Tests | |
| HHb, deoxyhemoglobin | O2Hb, oxyhemoglobin | 
| MetHb, methemoglobin | COHb, carboxyhemoglobin | 
| tHb, total hemoglobin | SO2%, oxygen saturation | 
| tBil, total bilirubin | HbF, fetal hemoglobin | 
Dimensions for Prime Plus, including CO-Oximetry and bidirectional connectivity:
     
Other Features 
    Full color, 10.1-inch touchscreen, multilingual, QC statistics, onboard data management, automatic sampler, integrated capillary adapter, optional barcode scanner, QC data storage, optional mobile cart with UPS 
  
  MicroSensor Card
    60 µL Sample Volume 
  
  Electrical Power Requirement 
  < 90 Watts 
  
  Acceptable Samples 
    Whole blood (heparinized), arterial, venous, mixed venous
    135 µL Sample draw requirement    
  
  Communication Protocols 
    ASTM, HL7, or POCT01-A2 connectivity formats
Certifications and Compliance: Nova Biomedical is certified to FDA Quality System Regulations and EN ISO 13485:2016 Complies to IVDD Tested according to: EN 61010-1:2010, EN 61010-2-101:2015, EN 60825-1/A1:2014.
Specifications current as of revision date.